A laboratory issues test reports under accreditation, and includes a unique serial number of the product tested in the test report. A statement as required by ISO 17025:2017 Cl. 7.8.2.1 l (a statement to the effect that the results relate only to the items tested) is not included. The laboratory states that the serial number is a unique identification of the sample tested.
- Is this in compliance with ISO/IEC 17025:2017?
When the answer is No:
- Is this acceptable when the laboratory agrees with the customer (in writing) that the statement can be waived? (also refer to ISO/IEC 17025:2017 Cl. 7.8.1.3)
No, the required statement is needed in all test reports to make clear to any user that the results cannot be extended to other items (i.e. cannot be interpreted as lot approval, product certification or inspection activities).